Page 15 - Water purification system for hemodialysis _ New guideline อ.กลศร
P. 15

When should chemical reanalysis of



                                    RO water be considered?










                    • Increased raw water conductivity or changing

                        source of raw water supply



                    • Increased RO water conductivity (>25 µS/cm)


                            • Especially in case of increased hardness of post-softener

                               water



                    • Decrease percent rejection of RO


                            • Rejection < 90 %

                            • Decrease > 10 % of previous value



                    • Patient(s) had suspicious clinical features of

                        toxicity from some chemical substance
   10   11   12   13   14   15   16   17   18   19   20