Page 97 - Water purification system for hemodialysis _ New guideline อ.กลศร
P. 97

Take home messages







              • In modern era of hemodialysis technology which high-flux &

                  super high-flux dialyzer and on-line producing substitution fluid

                  are used, the quality of water treatment system is crucial and

                  should be validated.



              • We can not prove the quality of dialysis fluid at real-time used

                  and the sterility of substitution fluid is beyond the lab sensitivity


                  to verify.


              • We have to use validation process to make sure that the quality

                  of dialysis fluid are within the required range at all time.



              • Using equipments according to validation process of

                  manufacturer, adequate monitoring & maintenance of water


                  treatment system, rapid corrective actions and revalidation after

                  correction are the principal task to achieve high fluid quality for

                  HD and ensuring patient safety.
   92   93   94   95   96   97   98