Page 114 - bukuneutron
P. 114
rouding the collimator.Inlet of collimator is located near
to the reactor core. The opening of inlet is large so
that the neutrons enter into the collimator can be max-
imised as much as possible.
Area for fully illuminated by neutron beam Wf is de-
termined by Equation 5.1. Where Ws is area of incom-
ing neutron beam, D is aperture size, Xap is the dis-
tance from source opening area to the aperture and L
is the length from aperture to detector plane [Grunauer,
2005].
Figure 5.5: Design construction parameter has the relationship
with the total illuminated area Wf of the collimator [Grunauer,
2005].
Wf = –D + Ws ( Xap + L) – Ws (5.1)
Xap
5.2.2 Image Detection and Recording
Conventional radiography used celulose film coated
with fluorescen materials was used to record the im-
101