Endotoxin Retentive Filter Performance









              • Logarithmic reduction value (LRV) for bacteria /endotoxin

                                           = log (A/B)
                                                  10
                      A = viable test bacteria /endotoxin in 1000 mL test solution

                      B = viable test bacteria /endotoxin in 1000 mL filtrate

              • JSDT standard (2011)


                             Standard dialysis fluid :          bacteria     < 100 CFU/mL       endotoxin  < 0.05 EU/mL

                             Ultrapure dialysis fluid :       bacteria     < 0.1 CFU/m          endotoxin  < 0.001 EU/mL

                                                                                    -6
                             Sterile (substitution) fluid : bacteria     < 10 CFU/mL       endotoxin  < 0.001 EU/mL





              • ETRF for obtaining ultrapure dialysis fluid

                             - LRV for bacteria at least 4  (usually  >7)

                             - LRV for endotoxin at least 2 (usually  >3)