Policies and Procedures on Infection Control


                     11.5.1   Administrative monitoring
                             •  Work practices must be supervised.
                             •  Written policy and procedures are strictly followed by all personnel
                                responsible for sterilizing and handling sterile supplies.
                             •  Policies and procedures pertain to the following:

                                -  Decontaminating, cleaning and terminally sterilizing.
                                -  Packaging and labeling
                                -  Loading and unloading the sterilizer
                                -  Operating the sterilizer
                                -  Monitoring and maintaining the record of each cycle
                                -  Adhering to safety precaution and preventive maintenance protocol
                                -  Transporting sterile packages to the sterile storage room.
                                -  Cart should be enclosed.
                                -  Storage of sterile items
                                -  Handling of sterile items
                                -  Tracking and recalling items if an item in a particular load is not safe for use.
                     11.5.2   Mechanical Monitoring
                              Routine maintenance to check efficiency and accuracy of autoclave consist of
                              the following:-
                             •  Dummy Run
                                A complete sterilizing cycle carried out with an empty chamber to get rid of
                                the remaining air in the chamber.
                             •  Recording Gauge
                                In the form of charts, printouts, or gauges, this reflects the current status of
                                cycle parameters (pressure, temperature and duration) during sterilization.
                                Gauge should be calibrated at regular intervals against standard instruments
                                by autoclave operator.
                             •  Leak Rate Test
                                Pre vacuum steam sterilizer must be tested at least once a week for the
                                rate of air leakage into the chamber during air removal and drying stages.
                             •  Thermocouple Test
                                To detect temperature achieved and maintained during sterilization stage.
                                This procedure should be done during commissioning, after major repair
                                and validation.
                             •  Bowie-Dick Test

                                Test packs are run daily, to monitor the function pre-vacuum sterilizers and
                                check the efficacy of vacuum system.


                 164  Ministry of Health Malaysia