Page 26 - LFC catalogue Metrology
P. 26
metrology
Flow Type Air Gage Display FT5000
The air micrometers are precision comparators which ingeniously apply the principles of
aerodynamics. They are widely used in fields which require machining of precision parts as
they have proved most eective in quality control and eciency improvement.
The FT-5000 series permits any number of air micrometers to be easily horizontally
stacked, which is very helpful for many quality control applications.
SPECIFICATIONS
FT-5100 FT-5200 FT-5500 FT-5000
Standard magnification (times) 1.000 2.000 5.000 10.000
Indication range (µm) 200 100 40 20
Effective indication range (µm) 150 70 30 15
Graduation (µm) 5 2 1 0.5
Equipment (µm) Max.3.0 Max.1.5 Max.0.8 Max.0.5
Repeatability (µm) Max.2.0 Max.1.0 Max.0.5 Max.0.3 FILTER UNIT :
Indication stability (µm) Max.2.0 Max.1.0 Max.0.5 Max.0.3 • Model FUA-703 (For single to three column use: 1 pc)
Response time (sec) Max.1.5 Max.1.5 Max.1.8 Max.2.0 • Model FUA-706 (For four to six column used: 1 pc)
Dimensions 110(W) x 210(D) x 501(H) mm
Analog Dial Type MD Series
The MD series dial pneumatic micrometers are precision comparators utilizing
the flow of high pressure air. Back pressure which is amplified in proportion to
a small clearance is indicated on the dial gauge. Unique adjustment over a
wide range. High pressure air blows off dust, oil, etc. to ensure accurate
measurement and quick response.
MD-14L The MD series contains the following two types: a single type which has a
single circuit indicating the measurement of one part and a duplex type which
has a duplex circuit indicating the measurement of two parts. Submicron
MD-24L
measurement is capable.
SPECIFICATIONS
FEATURES : Model number
Item MD-14L MD-24L
• The high pressure air type allows for a wide clearance between
the measuring probe and object, hence o�ers a wide measur- Number of needles indicator Single Twin
ing range and prolongs the probe life. Air supply (MPa) 0.4
• The precision regulator and amplifier relay assures accurate Indication range (µm) 200 100 50
2
1
5
measurement, stable indication, and very quick response. Graduation (µm) Max.2.5 Max.1.0 Max.0.5
The dual indicator has two pointers (black and red), facilitating Equipment (µm)
• the correlation of dimensions of two parts to be seen easily. Measuring clearance (mm) Max.0.12 Max.0.2 Max.0.25
Response time (sec) 0.5
Dimensions (mm) 190(W)x152(D)x177(H) 190(W)x152(D)x264(H)
Weight (Kg)
026 www.lfc.com.sg