Page 132 - AACN Essentials of Critical-Care Nursing Pocket Handbook, Second Edition
P. 132
Remeasure level of transducer air-reference and reposition at midchest level; rezero. Recheck zero and calibration of monitor. Check all connections. in proper sequence. Check air-reference level; reset at midchest and rezero. Review
Treatment Thoroughly flush transducers with IV solution; rezero and turn stopcocks Check position under x-ray and/or 119
Maintain transducer air-reference port at midchest level; rezero after patient Maintain transducer air-reference port at midchest level; recheck and rezero
Prevention position changes. Zero and calibrate monitor properly. Use Luer-Lock stopcocks. Use Luer-Lock stopcocks. Turn stopcocks in proper sequence when two pressures are measured on one transducer. after patient position changes. Establi
Inaccurate Pulmonary Artery Pressure Measurements (continued)
Incorrect air-reference level (above midchest level) Incorrect zeroing and calibration of monitor Loose connection Pressure trapped by improper sequence of stopcock operation Incorrect air-reference level (below midchest level) Migr
Cause
Problem Abnormally low or negative Abnormally high pressure Inappropriate pressure
pressures reading waveform
4.19