Page 136 - AACN Essentials of Critical-Care Nursing Pocket Handbook, Second Edition
P. 136

Press start button after injection completed to delay computer   Check for kinks, particularly at insertion site; straighten catheter;   123








                  Action  sampling time. Inject smoothly and quickly (10 mL in  4 seconds). Always check catheter patency by withdrawing, then flushing  proximal port before CO determinations.  reposition patient.




               Troubleshooting Problems with Thermodilution Cardiac Output Measurements (continued)







                     Magnetic interference producing numerous spikes in  CO curve Long lag time between injection and upstroke of   curve Uneven injection technique RA port partially occluded with clot  Catheter partially kinked





                  Cause














                  Problem  Irregular upslope of

                       CO curve
               4.20
   131   132   133   134   135   136   137   138   139   140   141