Page 135 - AACN Essentials of Critical-Care Nursing Pocket Handbook, Second Edition
P. 135

122
                   Inject exact volume to correspond to computation constant. Carefully remove all air bubbles from syringe. Verify PA waveform from distal lumen.  Pull catheter back.  Advance catheter.  Reposition patient. Rotate catheter to turn thermistor away







                Action                       Do not turn stopcock to reestablish IV infusion through proximal




             Troubleshooting Problems with Thermodilution Cardiac Output Measurements (continued)









                   Injectate volume less than designated amount  Catheter too distal (PAW)  RA port lies within sheath Thermistor against wall of PA   Fibrin covering thermistor   Incorrect  variables  Right-to-left shunt (VSD) Severe tricuspid regurgitation Incor



                Cause














                Problem  Cardiac output values   higher than expected