Page 138 - Clinical Application of Mechanical Ventilation
P. 138
104 Chapter 4
TABLE 4-6 Characteristics of the Pressure Support Ventilation Mode
Characteristic Description
Type of breath Pressure-supported breaths are considered spontaneous. (Note:
Pressure support may be applied in any mode that permits
spontaneous breathing such as SIMV.)
Triggering mechanism Pressure-supported breaths are patient-triggered.
Cycling mechanism Pressure-supported breaths are technically flow-cycled by a minimum
spontaneous inspiratory flow threshold. This minimum inspiratory flow
is controlled entirely by the patient’s spontaneous inspiratory flow
demand.
Notes: The tidal volume delivered by a pressure-supported breath is influenced both by the pressure support level (cm H 2 O) and the patient’s spontaneous
inspiratory flow demand. The inspiratory time of the pressure-supported breath is also completely controlled by the patient’s spontaneous inspiratory
flow demand.
© Cengage Learning 2014
ADAPTIVE SUPPORT VENTILATION (ASV)
The adaptive support ventilation (ASV) is a dual control mode that provides
adaptive support ventilation
(ASV): A mode of ventilation that a mandatory minute ventilation. The ventilator measures the dynamic compli-
changes the number of mandatory ance and expiratory time constant to adjust the mechanical tidal volume and
breaths and pressure support level
according to the patient’s breath- frequency for a target minute ventilation. The optimal tidal volume is calcu-
ing pattern. lated by dividing the minute ventilation by the optimal frequency in terms of
the lowest work of breathing. Basically, ASV uses the Otis Equation to calcu-
late the optimal frequency that corresponds with the lowest work of breathing
(Chen et al., 2008; Otis et al., 1950). Figure 4-9 shows the selection of optimal
frequency based on the lowest total work of breathing (WOB tot) (Lotti et al.,
2002).
With ASV mode, the therapist inputs the patient’s body weight and the desired
percent minute volume. The body weight is used to estimate the deadspace volume
and to calculate the alveolar volume. For an estimated minute ventilation require-
ment for a patient, the ventilator uses predetermined settings of 100 mL/min/kg for
adults and 200 mL/min/kg for children. The therapist may select the percent min-
ute volume, ranging from 20% to 200% of the predetermined adult or child setting.
For example, if 160% is selected for an adult, the minute ventilation delivered by
the ventilator will be about 160 mL/min/kg.
Once the target minute ventilation is set, the ventilator uses test breaths to mea-
sure the system compliance, airway resistance, and any intrinsic PEEP. Following
determination of these variables, the ventilator selects and provides the frequency,
inspiratory time, I:E ratio, and high pressure limit for mandatory and spontaneous
breaths.
Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).
Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.

