Page 587 - Clinical Application of Mechanical Ventilation
P. 587

Neonatal Mechanical Ventilation  553


                                             respiratory acidosis (pH less than 7.20), or unsatisfactory PaO  on .50% of oxy-
                                                                                                   2
                                             gen with N-CPAP, intubation and surfactant therapy should be considered.


                        BASIC PRINCIPLES OF NEONATAL VENTILATION



                                             The primary mode of neonatal mechanical ventilation is pressure-controlled venti-
                                             lation. The ventilator generates a sufficient flow and delivers variable tidal volumes
                                             by the preset pressure. Alternatively, ventilation can be achieved by using a preset
                                             volume (volume-controlled ventilation).

                                             Pressure-Controlled Ventilation


                                             In pressure-controlled ventilation, a preset peak inspiratory pressure (PIP) is used
                        pressure-controlled ventilation:
                        A preset pressure is used to deliver   to deliver the volume. Thus, pressure is constant and volume is variable, depending on
                        tidal volumes; the delivered volumes   the compliance and airflow resistance characteristics of the infant. A decreasing com-
                        are variable.
                                             pliance or an increasing airflow resistance requires a higher pressure to maintain the
                                             same tidal volume. With the use of pressure-controlled ventilation, frequency would
                                             more likely be adjusted to maintain minute ventilation, since the strategy of setting a
                                             PIP is to protect the lung from excessive airway pressures. As the patient’s condition
                                             improves (increased compliance or decreased airflow resistance), the pressure must be
                            Following surfactant   decreased to avoid excessive volume and pressure. An example of this observation is
                          replacement, the PIP must
                          be monitored and reduced   the successful therapeutic response to surfactant replacement. The lung compliance
                          accordingly to avoid overex-
                          pansion of the lungs.  of the infant increases dramatically and rapidly soon after surfactant administration.
                                             The PIP must be reduced accordingly to avoid overexpansion of the lungs.

                                             Volume-Controlled Ventilation


                                             Volume-controlled ventilation has gained popularity because the volume is preset
                        volume-controlled ventilation:
                        A desired tidal volume is preset;   using variable pressures. This strategy reduces the incidence of overexpansion of
                        the pressure needed to deliver the   the lungs. For extremely low-birth-weight infants, some ventilators (e.g., Drager
                        volume is variable.
                                             Babylog) can provide a tidal volume as low as 2 mL. This is an important feature
                                             when trying to maintain a range of 3–7 mL/kg for extremely low-birth-weight
                                             infants (e.g., as low as 500 g birth weight). Typically, an initial tidal volume of
                                             5 mL/kg is used to ventilate these infants, and the volume may be titrated up or
                                             down to meet the infant’s needs. It is important to note that although volume
                                             remains constant, pressure will vary, which may result in excess pressures being
                                             delivered to the infant as a result of a decrease in compliance or increase in airway
                                             resistance. Careful setting of peak pressure alarms will aid in preventing excessive
                                             pressures being delivered in volume-controlled ventilation.

                                             Ventilator Circuits and Humidifiers


                                             During mechanical ventilation, some of the ventilator volume is “lost” within the
                                             circuit and humidifier and is not delivered to the patient. This wasted volume is






                        Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).
                      Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.
   582   583   584   585   586   587   588   589   590   591   592