Page 591 - Clinical Application of Mechanical Ventilation
P. 591

Neonatal Mechanical Ventilation  557



                          TABLE 17-6 Initial Settings for Neonatal Mechanical Ventilation Based on Compliance

                          Parameter     Normal Compliance             Low Compliance

                          PIP           15 to 20 cm H O               20 to 30 cm H O
                                                                                  2
                                                     2
                          PEEP          3 to 5 cm H O                 Up to 8 cm H O
                                                  2
                                                                                 2
                          V *           4 to 8 mL/kg                  6 to 10 mL/kg
                            T
                          Frequency     25 to 40/min                  Up to 150/min (esp. with air leak)

                          Flow Rate     6 to 8 L/min                  6 to 8 L/min
                          I Time        0.3 to 0.5 sec                Change according to frequency to maintain an
                                                                        I:E ratio of 1:1
                          I:E Ratio     1:1.5 to 1:2                  At least 1:1 (to prevent inverse I:E ratio)

                          F O 2         Set to keep patient pink with   Set to keep patient pink with SpO  from 90%
                                                                                                    2
                           I
                                          SpO  from 90% to 95%          to 95%
                                              2
                                                                      Use with appropriate PEEP level if necessary
                        *On	ventilators	that	are	capable	of	providing	a	preset	tidal	volume	(volume-controlled	ventilation).	(Data	from	Bandy	et	al.,	1992;	Goldsmith
                        et	al.,	1993;	Deakins	et	al.,	2001;	Walsh	et	al.,	2010.)
                        © Cengage Learning 2014

                                             The frequency on many neonatal ventilators is controlled by adjusting the expira-
                                             tory time after setting the flow and inspiratory time. Generally the frequency is
                                             set from 40 to 60/min for premature infants and 25 to 40/min for term infants
                                             (Table 17-7). The frequency is dependent on the ventilatory requirement and dis-
                                             ease process. The inspiratory time is usually set at 0.3 to 0.5 sec, depending on
                                             gestational age, with a shorter inspiratory time when the frequency is high or in the
                                             presence of air trapping. Using the neonate’s body weight as a guide, 5 mL/kg is the
                                             suggested initial tidal volume with a range of 3–7 mL/kg. The PEEP is set between
                                             4 and 8 cm H O depending on the oxygenation status of the infant. A higher PEEP
                                                         2
                                             is used initially if the infant’s F O  requirement is high, if there is evidence of severe
                                                                       I
                                                                          2


                          TABLE 17-7 Initial Settings for Neonatal Mechanical Ventilation Based on Weight

                          Parameter          Very Low Birth Weight   Low Birth Weight        Term

                          Frequency          40 to 60/min            40 to 60/min            25 to 40/min
                          Inspiratory time   0.3 to 0.5 sec          0.3 to 0.5 sec          0.3 to 0.5 sec

                          Volume             3 to 7 mL/kg starting at   3 to 7 mL/kg starting   3 to 7 mL/kg starting at
                                              5 mL/kg                  at 5 mL/kg             5 mL/kg
                          PEEP               4 to 8 cm H O           4 to 8 cm H O           4 to 8 cm H O
                                                                                                       2
                                                       2
                                                                                2
                        © Cengage Learning 2014




                        Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).