SECTION 5.0 – ARCHITECTURAL COMPONENTS, MATERIALS AND FINISHES   SECTION 5.0 – ARCHITECTURAL COMPONENTS, MATERIALS AND FINISHES



 j)   All differ, master, grand master keys are to be sealed, labelled, and   g)  The  minimum  size  of  closer  must  comply  with  EN  1154  (500,000
 handed  to  the  Client  upon  completing  the  project  in  a  properly   testings), or any accredited recognised international standard size 3 to
 organised manner to the PD’s satisfaction and concurrence.    5 depending on the door weight and width and shall have five (5) years
                                     warranty. The door closers can hold up more than 100 kg door.
 k)  All lock strike plates shall be supplied with a box and shall have a lip of
 enough length to protect the door trim and jamb. Mortise backset shall   h)  For areas involving special interiors such as the auditorium, Royal
 be not less than 55 mm (single throw) and  60 mm (double throw).   Ward, VIP Room, and other specified areas, the door closer shall come
                                     with a sliding arm with  two pressure  relief valves, including latching
 l)   Tubular knob handles shall not be acceptable for fire doors.    speed and closing speed for better function and aesthetics.

                                  i)  Toilet cubicle doors shall be fitted with indicator locks.
 5.11.10  Hinges

    All hinges supplied shall be from the same manufacturer and comply with   5.11.12  Floor spring
 the latest JKR Standard Specification for Building Work. Unless otherwise
 specified, stainless steel grade 304 butt weld hinges of not less than three   a)  All floor springs shall be provided with a pressure relief valve to prevent
 (3) hinges per door leaf shall be provided, depending on the door weight   overloading and oil leakage. The floor spring shall be supplied with an
 and width.                           adjustable hold-open function and a five (5) years warranty.
                                  b)  Heavy-duty floor spring shall be incorporated with a hold-open function,
                                      and a minimum door weight of 120 kg should be allowed.
 5.11.11  Door Closers and Door Stoppers
                                  c)  Stainless steel of minimum grade 304 floor spring cover plates shall be
 a)  Door closers and door stoppers shall be of the approved locally   installed correctly to avoid/reduce dust collection in the cement case
 manufactured type and shall be properly installed and fastened where   and mounting.
 they shall not obstruct any equipment, furniture and/or services.
                        5.11.13  The  successful  tenderer  shall  provide  customised  computer-generated
 b)  All doors shall be fitted with aluminium alloy door closers except toilet   schedules with the following information:
 cubicle doors and sliding doors, and shall have a hold-open function
 except for fire doors.           a)  Consolidated break down of hardware sets.

 c)  Door closer elements such as spring/power setting, back-check, sweep   b)  Consolidated door schedule by the floor and by building.
 speed, and latch speed valve shall be adjustable.
                                  c)  Consolidated door hardware sets by items.
 d)  Toilet cubicle doors shall be fitted with one (1) stainless steel hook.
                                  d)  Consolidated door schedule with master key reference.
 e)  When the operating force needed to open the door is greater than     25
 N  for  PWD  toilet  doors,  an  automatic  opening  door  is  required.   e)  Consolidated door schedule with grand master key reference.
 Preferably automatic doors or controlled door closing devices with a
 hold-open  device.  An  alternative  option  is  using  dual-powered-  f)  Consolidated door schedule with great grand master key reference.
 controlled door locking devices with electro-magnetic retention for
 higher power spring.             g)  Consolidated door schedule with key alike

 f)  All  door  closers  shall  have  two  pressure  relief  valves  to  prevent  oil   h)  Door type, fire door, and timber flush door.
 leakage and damage to the back-check valve. It must be completely
 reversible, have a back feature, and have a warranty of less than five   i)  Any other schedule that may be needed and requested by the PD.
 (5) years from the manufacturer.