contains the He3 detector is installed in the middle of
the sphere. The principle of neutron using He3 or BF3
is already discussed in Chapter 1.The detector is con-
nected to the readout electronic and display as it is
shown in Figure 6.2 and Figure ??.

Figure 6.2: Scalar rate meter readout is connected to
the detector He3 or BF3 locate inside the polyethylen
sphere of Ludlum detector type 12-4.

    In general, survey meter gives data on the dose
rate and its required to convert into the dose equva-
lent (H) per neutron fluence (i.e., rem per neutron). It
is varies with the neutron energy because both the ab-
sorbed dose (D) and the quality factor (Q,Wx) vary with
the neutron energy. Therefore, the dose equivalent (H)
or the effective dose can be calculated using Equation
6.4.Where, the value of Q and Wx is given in Table 6.1.