Page 426 - Clinical Application of Mechanical Ventilation
P. 426

392    Chapter 12


                                            distress	possibly	due	to	conditions	such	as	hypoxia,	pain,	anxiety,	inadequate	inspiratory
                                            flow	or	pressure	support.	Tachypnia	shortens	the	expiratory	time,	increases	the	mean
                                            airway	pressure,	and	alters	the	ventilation/perfusion	relationship.	For	these	reasons,	per-
                                            sistent	tachypnia	must	be	investigated	and	corrected.	Triggering	of	the	high	frequency
                                            alarm	may	be	corrected	by	simple	steps	such	as	endotracheal	suctioning,	increasing	the
                                            F O ,	peak	flow	or	pressure	support	setting.	The	high	frequency	alarm	limit	must	not	be
                                               2
                                             I
                                            increased	without	clear	justification	(e.g.,	reversal	of	sedation	or	anesthesia).
                                             Another	cause	of	the	high	frequency	alarm	is	an	inappropriate	sensitivity	set-
                                            ting.	When	this	control	is	set	excessively	sensitive	to	the	patient’s	inspiratory	effort,
                                            minimum	inspiratory	efforts	or	movements	will	cause	the	ventilator	to	initiate	auto-
                                            triggering	and	increase	in	total	frequency.

                                            Apnea/Low Frequency Alarm


                                            The	apnea/low	frequency	limit	is	set	to	ensure	that	a	minimum	number	of	breaths
                                            is	delivered	to	the	patient.
                                             The	apnea	or	low	frequency	alarm	is	triggered	when	the	total	frequency	drops
                          Disconnection of the   below	the	low	frequency	limit.	Disconnection	of	the	ventilator	circuit	from	the
                        ventilator circuit from the
                        patient’s endotracheal tube is   patient’s	endotracheal	tube	is	the	most	frequent	trigger	of	the	apnea	alarm,	since	the
                        the most frequent trigger of   ventilator	cannot	sense	any	air	movement	(respiratory	effort)	from	a	disconnected
                        the apnea alarm.
                                            circuit.	Other	triggers	of	the	apnea/low	frequency	alarm	include	a	patient	under
                                            respiratory	depressants	or	muscle-paralyzing	agents,	conditions	of	respiratory	center
                                            dysfunction,	and	respiratory	muscle	fatigue.
                                             Some	ventilators	merely	alert	the	practitioner	that	the	patient	is	having	periods
                                            of	apnea;	the	practitioner	 must	increase	ventilation	to	 alleviate	the	situation.
                                            Most	 ventilators	 switch	 to	 a	 backup	 ventilation	 mode	 until	 the	 problem	 is
                                            corrected.

                                            High PEEP Alarm


                                            The	high	PEEP	limit	is	set	to	prevent	excessive	PEEP	imposed	on	the	patient.	The
                                            alarm	is	triggered	when	the	actual	PEEP	exceeds	the	preset	PEEP	limit.	Auto-PEEP
                                            may	occur	in	conditions	of	air	trapping,	insufficient	inspiratory	flow	(long	I-time),
                                            or	insufficient	expiratory	time	(short	E-time).
                                             Air	 trapping	 may	 be	 reduced	 by	 decreasing	 the	 ventilator	 tidal	 volume	 and
                                            frequency,	and	by	using	bronchodilators	in	patients	with	reversible	airway	obstruc-
                                            tion.	Increasing	the	inspiratory	peak	flow	provides	a	shorter	I-time	and	a	longer
                                            E-time.	More	time	for	exhalation	helps	to	reduce	air	trapping.

                                            Low PEEP Alarm


                                            The	low	PEEP	limit	is	set	to	ensure	that	the	preselected	PEEP	is	delivered	to	the
                                            patient.	The	alarm	is	triggered	when	the	actual	PEEP	drops	below	the	preset	low
                                            PEEP	limit.	Failure	of	the	ventilator	circuit	to	hold	the	PEEP	is	usually	due	to
                                            leakage	in	the	circuit	or	ET	tube	cuff.






                        Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).
                      Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.
   421   422   423   424   425   426   427   428   429   430   431