Page 431 - Clinical Application of Mechanical Ventilation
P. 431

Management of Mechanical Ventilation  397


                                             for	the	heated	humidifier	should	be	placed	inside	the	inspiratory	limb	of	the	venti-
                                             lator	circuit	as	close	to	the	patient	as	possible.	Since	water	vapor	saturation	depends
                                             on	the	water	content	as	well	as	the	temperature,	the	temperature	setting	should	be
                                             adjusted	for	a	distal	temperature	reading	of	37°C.	This	ensures	proper	temperature
                                             and	humidification	to	the	patient	(Burton	et	al.,	1997).

                                             Frequency of Circuit Change


                                             Ventilator	circuits	should	not	be	changed	routinely	for	infection	control	purposes.
                                             The	maximum	duration	of	time	that	circuits	can	be	used	safely	is	unknown	(Hess
                                             et	al.,	2003).	For	circuits	with	a	humidifier	or	HME,	they	should	be	changed	only
                                             when	 visibly	 soiled	 (Tablan	 et	 al.,	 2004).	 Studies	 have	 shown	 that	 the	 optimal
                            The optimal interval for   interval	for	ventilator	circuit	change	is	once	per	week	(Fink,	1998;	Kotilainen,	1997;
                          ventilator circuit change is
                          once per week.     Long	et	al.,	1996;	Stamm,	1998).	When	compared	to	more	frequent	circuit	changes,
                                             weekly	circuit	change	does	not	increase	the	incidence	of	nosocomial	infection,	in-
                                             cluding	ventilator-associated	pneumonia.	Weekly	change	also	saves	manpower	and
                                             reduces	the	direct	replacement	cost	for	new	ventilator	circuits	(Kotilainen,	1997).


                        CARE OF THE ARTIFICIAL AIRWAY



                                             Supplemental	humidity	must	be	provided	during	mechanical	ventilation,	because
                            Patency of the ET tube   the	endotracheal	(ET)	tube	does	not	receive	humidification	normally	provided	by
                          can only be ensured with
                          adequate humidification and   the	upper	airway.	In	addition,	secretions	must	be	removed	by	suctioning,	if	nec-
                          prompt removal of retained   essary,	because	the	ET	tube	and	the	ventilator	circuit	are	a	closed	system.	If	not
                          secretions.
                                             removed,	any	secretions	coughed	up	by	the	patient	are	likely	to	stay	in	the	ET	tube.
                                             Patency	of	the	ET	tube	can	only	be	ensured	with	adequate	humidification	and
                                             prompt	removal	of	retained	secretions.

                                             Patency of the Endotracheal Tube


                                             In	mechanical	ventilation,	the	primary	purpose	of	an	ET	tube	is	to	protect	the
                                             airway	and	to	provide	airflow	to	the	lungs.	Since	airflow	resistance	is	inversely
                                             related	to	the	diameter	of	the	tube,	small	tubes	cause	a	tremendous	increase	in
                                             the	work	of	breathing.	In	order	to	maximize	airflow,	the	largest	ET	tube	that	is
                                             appropriate	to	the	patient	should	be	used.	Mucus	in	the	ET	tube	should	also	be
                                             removed	frequently	in	order	to	minimize	airflow	obstruction	created	by	retained
                                             secretions.
                                               Poiseuille’s	Law	shows	that	when	the	radius	of	an	airway	is	reduced	by	half,	the
                                             driving	pressure	(work	of	breathing)	must	be	increased	16	times	in	order	to	main-
                                             tain	the	same	flow	rate.	An	obstructed	airway	hinders	not	only	mechanical	ventila-
                                             tion,	but	spontaneous	ventilation	as	well.	Airway	management	should	always	be	an
                                             integral	part	of	mechancial	ventilation.

                                                                                       Flow
                                                                      Pressure	change =
                                                                                        r 4






                        Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).
                      Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.
   426   427   428   429   430   431   432   433   434   435   436