APPENDIX G  ■  The Microscope                         723






                   Step          Procedural Steps



                   16            Following examination of a specimen slide, remove the slide and clean the microscope by wiping off the objective

                                 lens with lens paper. Be careful to clean the oil immersion lens last.

                   17            Return the low power (10×) objective to the viewing position and completely decrease the working distance by lowering

                                 the objective to the lowest possible position.


                   18            Cover the microscope and return to storage, if it is not stored on the laboratory benchtop.







                   TIPS FOR CLEANING THE MICROSCOPE                                                                            a circular motion, beginning at the center an   moving out-

                                                                                                                               war  . Repeat with new   ampene   lens paper as necessary.

                   Cleaning the Exterior                                                                                       Finally, blot   ry with clean lens paper. Do not rub; this may

                                                                                                                               scratch the sur ace o  the lens.
                   T e sur ace o  most microscopes may be kept clean by wash-                                                       Lenses  shoul    never  be  touche    with  the  f ngers.

                   ing it with a neutral soap an   water.   o clean the metal an                                               Objectives  must  not  be  taken  apart  because  even  a  slight

                   enamel,  a  gauze  or  so    cloth  shoul    be  moistene    with                                           alteration o  the lens setting may ruin the objective. Merely

                   the cleaning agent an   rubbe   over the sur ace with a cir-                                                clean the outer sur ace o  the lens as   escribe  . An espe-

                   cular  motion.  T e  sur ace  shoul    be    rie    imme  iately                                            cially   irty objective may be unscrewe    rom the nosepiece

                   with a Kimwipe or clean,   ry piece o  gauze. Kimwipes or                                                   then hel   upsi  e   own an   checke    or cleanliness by using

                   gauze shoul   never be use   to clean the optical parts o  the                                              the ocular (remove    rom the bo  y tube) as a magni ying

                   microscope.
                                                                                                                               glass. Dust or lint can also be remove    rom the rear lens o

                                                                                                                               the objective by blowing it away with an air syringe. Such
                   Cleaning Optical Lenses: General                                                                            removal o  the objective  rom the nosepiece is not a routine

                   Comments                                                                                                    cleaning proce  ure. T e f nal step a  er using the microscope



                   T e glass sur aces o  the ocular, the objectives, an   the con-                                             shoul   always be to wipe o   all objectives with clean lens

                     enser are han  -groun   optical lenses. T ese lenses must be                                              paper.

                   kept meticulously clean. Optical glass is so  er than or  inary

                   glass an   shoul   never be cleane   with paper tissue or gauze;                                            Cleaning the Ocular

                   these materials will scratch the lens.   o clean the lenses o  the                                          Te ocular (eyepiece) is susceptible to   irt because o  its loca-

                   microscope, use lens paper.                                                                                 tion on the microscope an   contact with the observer’s eye.

                        Be ore polishing with lens paper, take care that nothing                                               Mascara presents a constant cleaning problem. Dust can be

                   is present that will scratch the optical glass in the polishing                                             remove    rom the lens o  the ocular with an air syringe or

                   process. Such potentially abrasive   irt,   ust, or lint can easily                                         camel’s hair brush; air is probably easier to use an   more

                   be blown away be ore polishing.                                                                             e  cient. T e lens shoul   then be polishe   with lens paper.


                                                                                                                               T e ocular can be checke    or a    itional   irt by hol  ing it
                   Cleaning the Objectives                                                                                     up to a light an   looking through it. When looking into the


                   Oil must be remove    rom the oil-immersion (100×) objec-                                                   microscope,   irt on any part o  the ocular will rotate with the

                   tive imme  iately a  er use by wiping with clean lens paper.                                                ocular when it is turne  . T e ocular shoul   not be remove

                   I  not remove  , oil may seep insi  e the lens or   ry on the                                                or more than a  ew minutes;   ust can collect in the bo  y tube

                   outsi  e sur ace o  the objective. T e high-  ry (40×) objective                                            an   settle on the rear lens o  an objective.

                   shoul   never be use   with oil, but i  this or any other objec-

                   tive or microscope part comes into contact with oil, it shoul                                               Cleaning the Condenser

                   be cleane   imme  iately. I  a lens is especially   irty, it may                                            T e light source an   con  enser shoul   also be  ree o    ust,

                   be cleane   with a small amount o  commercial lens cleaner,                                                 lint, an     irt. First, blow away the   ust with an air syringe

                   methanol, or manu acturer-recommen  e   solution applie                                                     or camel’s hair brush, then polish the light source an   con-

                   to lens paper then wipe   across the sur ace.                                                                enser with lens paper. It may be necessary to clean these

                          o clean the oil-immersion lens properly, f rst lower the                                             parts  urther with lens paper moistene   with a commercial

                   stage, then rotate the objective to the  ront an   wipe gently                                              lens cleaner or methanol be ore polishing them with lens

                   with clean lens paper. Clean o   the immersion oil with lens                                                paper.

                   paper   ampene   with special lens cleaner.

                        Alternatively,  the  cleaning  agent  may  be  applie    to  a                                         Cleaning the Stage and Adjustment Knobs

                   woo  en applicator stick wrappe   with cotton or lens paper

                   an   moistene   with the cleaning agent. Do not use a plastic                                               T e stage o  the microscope shoul   be cleane   a  er each use

                   applicator stick; it will be   issolve   by the solvent, ruining                                            by wiping with gauze or a tissue. A  er it has been cleane

                   the objective. Apply the cleaning agent by blotting an   using                                              thoroughly, the stage shoul   be wipe     ry.