Page 128 - Clinical Application of Mechanical Ventilation
P. 128
94 Chapter 4
TABLE 4-2 Characteristics of the Control Mode
Characteristic Description
Type of breath Each breath delivers a mechanical tidal
volume.
Triggering mechanism Every breath in the control mode is
time-triggered.
Cycling mechanism Inspiration is terminated by the delivery
of a preset tidal volume (volume-cycled).
© Cengage Learning 2014
Complications of Control Mode
Since the patient’s spontaneous respiratory drive will have been blunted with seda-
In a sedated or apneic tion and neuromuscular block in the control mode, the patient is totally dependent
patient, the primary hazard
of the control mode is the on the ventilator for ventilation and oxygenation. As a result, the primary hazard
potential for apnea and associated with the control mode is the potential for apnea and hypoxia if the pa-
hypoxia if the patient should
become disconnected from tient should become accidentally disconnected from the ventilator or the ventilator
the ventilator or the ventilator should stop to operate due to mechanical or electrical failure.
should fail to operate.
Another physiologic problem with the control mode is the rapid disuse atrophy of
diaphragm fibers. One study shows that a combination of 18 to 69 hours of complete
diaphragmatic inactivity during mechanical ventilation results in marked atrophy of
human diaphragm myofibers (Levine et al., 2008). Furthermore, prolonged mechani-
cal ventilation leads to diaphragmatic oxidative injury, elevated proteolysis, and reduced
function of the diaphragm (Betters et al., 2004). Because of the patient’s dependence on
the ventilator, the most important alarms in the control mode become those that alert
the caregiver of any interruption in the patient’s ventilation. The important ventilator
alarms include the low exhaled volume alarm and the low inspiratory pressure alarm.
Table 4-2 summarizes the major characteristics of the control mode.
ASSIST/CONTROL (AC)
With the assist/control (AC) mode, the patient may increase the ventilator frequency
(assist) in addition to the preset mechanical frequency (control). Each control breath
provides the patient with a preset, ventilator-delivered tidal volume. Each assist
breath also results in a preset, ventilator-delivered tidal volume. The assist control
mode does not allow the patient to take spontaneous breaths (Figure 4-5).
Assist Control Triggering Mechanism
The mandatory mechanical breaths may be either patient-triggered by the patient’s
spontaneous inspiratory efforts (assist) or time-triggered by a preset frequency
Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).
Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.

