Page 315 - Clinical Application of Mechanical Ventilation
P. 315

Hemodynamic Monitoring  281



                          TABLE 10-5 Potential Problems with Arterial Catheter

                          Factors                                       Problem

                          Air bubbles in tubing                         Dampens the pressure signal
                          Loose tubing connections
                          Transducer and catheter placed higher than    Measurement lower than actual
                            measurement site

                          Transducer and catheter placed lower than     Measurement higher than actual
                            measurement site

                          Inadequate pressure applied to the flush      Backup of blood in the tubing
                            solution bag

                          Blood clot at catheter tip, catheter tip blocked   Inaccurate reading, signal interference
                            by wall of artery
                        © Cengage Learning 2014



                                             Potential Problems with Arterial Catheter


                                             Air  bubbles  and  loose  tubing  connections  can  “dampen”  the  pressure  signal.
                                             Improper  leveling  of  the  transducer  and  catheter  can  cause  false  high  or  false
                                             low readings. Inadequate pressure applied to the heparin solution bag can result
                                             in backup of blood in the tubing when the arterial pressure becomes higher than
                                             the heparin line pressure. Clotting of blood at the catheter tip or blockage of the
                                             catheter tip by the wall of the artery can interfere with the hemodynamic signal. The
                                             potential problems that are related to the arterial catheter are shown in Table 10-5.
                                               Most intensive care units have standard procedures in place to minimize such
                                             problems.  Careful  adherence  to  proper  setup  and  calibration  of  hemodynamic
                                             monitoring equipment are essential.



                        CENTRAL VENOUS CATHETER



                                             The central venous pressure (CVP) can be monitored through a central venous
                                             catheter placed either in the superior vena cava near the right atrium or in the right
                                             atrium. The pressure measured in the right atrium is right atrial pressure (RAP) but
                                             it is commonly called CVP. The RAP can also be monitored via the proximal port
                                             of a pulmonary artery catheter.
                                               The primary use CVP in hemodynamic monitoring is to measure the filling pres-
                            CVP measures the filling
                          pressures in the right heart   sures in the right heart. The CVP is helpful in assessing fluid status and right heart
                          and assesses the systemic   function. However, it is often late to reflect changes in the left heart. The central
                          fluid status and right heart
                          function.          venous catheter can also be used to collect “mixed” venous blood samples and for
                                             administration of medications and fluids. (Note: A true mixed venous blood sample






                        Copyright 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part. Due to electronic rights, some third party content may be suppressed from the eBook and/or eChapter(s).
                      Editorial review has deemed that any suppressed content does not materially affect the overall learning experience. Cengage Learning reserves the right to remove additional content at any time if subsequent rights restrictions require it.
   310   311   312   313   314   315   316   317   318   319   320