Page 219 - ACCCN's Critical Care Nursing
P. 219

196  P R I N C I P L E S   A N D   P R A C T I C E   O F   C R I T I C A L   C A R E

         should reflect how much information is required to opti-  monitor should display zero (0 mmHg), as this equates
         mise the patient’s condition, and how precisely the data   to current atmospheric pressure (760 mmHg at sea level).
         are to be recorded. As Pinsky argues, a great deal of infor-  With  the  improved  quality  of  transducers,  repeated
         mation is generated by this form of monitoring, and yet   zeroing is not necessary, as once zeroed, the drift from
                                            24
                                                                                      29
         little  of  this  is  actually  used  clinically.   Consequently,   the  baseline  is  minimal.   Some  critical  care  units,
         monitors are not substitutes for careful examination and   however,  continue  to  recalibrate  transducer(s)  at  the
         do not replace the clinician. The accuracy of the values   beginning of each clinical shift.
         obtained and a nurse’s ability to interpret the data and   Fast-flush square wave testing, or dynamic response mea-
         choose  an  appropriate  intervention  directly  affect  the   surement,  is a way of checking the dynamic response of
                                                                       29
         patient’s condition and outcome. 25
                                                              the monitor to signals from the blood vessel. It is also a
         PRINCIPLES OF HAEMODYNAMIC                           check on the accuracy of the subsequent haemodynamic
                                                              pressure values. The fast-flush device within the system,
         MONITORING                                           when  triggered  and  released,  exposes  the  transducer  to
         A  number  of  key  principles  need  to  be  understood  in   the amount of pressure in the flush solution bag (usually
         relation  to  invasive  haemodynamic  monitoring  of  the   300 mmHg). The pressure waveform on the monitor will
         critically ill patients. These include haemodynamic accu-  show  a  rapid  rise  in  pressure,  which  then  squares
         racy,  the  ability  to  trend  data  and  the  maintenance  of   off  before  the  pressure  drops  back  to  the  baseline  (see
         minimum standards. These are reviewed below.         Figure 9.17).

         Haemodynamic Accuracy                                Interpretation of the square wave testing is essential; the
                                                              clinician  must  observe  the  speed  with  which  the  wave
         Accuracy  of  the  value  obtained  from  haemodynamic   returns to the baseline as well as the pattern produced.
         monitoring is essential, as it directly affects the patient’s   One to three rapid oscillations should occur immediately
         condition. 26,27  Electronic equipment for this purpose has   after  the  square  wave,  before  the  monitored  waveform
         four components (see Figure 9.16):                   resumes.  The  distance  between  these  rapid  oscillations
                                                                                                 29
            1.  an  invasive  catheter  attached  to  high-pressure   should  not  exceed  1 mm  or  0.04 sec.   Absence,  or  a
               tubing                                         reduction, of these rapid oscillations, or a ‘square wave’
            2.  a transducer to detect physiological activity  with rounded corners, indicates that the pressure moni-
            3.  a flush system                                toring system is overdamped; in other words its respon-
            4.  a  recording  device,  incorporating  an  amplifier  to   siveness to monitored pressures and waveforms is reduced
               increase  the  size  of  the  signal,  to  display   (see Figure 9.18). An underdamped monitoring system
               information.                                   will produce more rapid oscillations after the square wave
                                                              than usual.
         High-pressure  (non-distensible)  tubing  reduces  distor-
         tion  of  the  signal  produced  between  the  intravascular   Data Trends
         device and the transducer; the pressure is then converted   The ability to trend data via a monitor or a clinical infor-
         into electrical energy (a waveform). Fluid (0.9% sodium   mation system is essential for critical care practice. Current
         chloride) is routinely used to maintain line patency using   monitoring systems used in Australia and New Zealand
         a continuous pressure system; the pressure of the flush   can retain data for a period of time, produce trend graphs,
         system  fluid  bag  should  be  maintained  at  300 mmHg,   and link to other devices to allow review of data from
         which normally delivers a continual flow of 3 mL/h.
                                                              locations  other  than  the  immediate  bedside.  The  data
         Accuracy is dependent on levelling the transducer to the   trends can be used to assess the progression of a patient’s
         appropriate level (and altering this level with changes in   clinical condition and monitor the patient’s response to
         patient position as appropriate), then zeroing the trans-  treatment.
         ducer in the pressure monitoring system to atmospheric
         pressure  (called  calibration)  as  well  as  evaluating  the   Haemodynamic Monitoring Standards
         response  of  the  system  by  fast-flush  wave  testing.    There are stated minimum standards for critical care units
         The transducer must be levelled to the reference point of   in Australia and New Zealand. 30,31  The standards require
         the phlebostatic axis, at the intersection of the 4th inter-  that  patient  monitoring  include  circulation,  respiration
         costal space and the midthoracic anterior-posterior diam-  and oxygenation, with the following essential equipment
                                     27
         eter  (not  the  midaxillary  line).   Error  in  measurement   available  for  every  patient:  an  ECG  that  facilitates  con-
         can occur if the transducer is placed above or below the   tinual cardiac monitoring; a mechanical ventilator, pulse
         phlebostatic  axis. 26,27   Measurements  taken  when  the   oximeter; and other equipment available where necessary
         patient  is  in  the  lateral  position  are  not  considered  as   to measure intra-arterial and pulmonary pressures, cardiac
         accurate as those taken when the patient is lying supine   output, inspiratory pressure and airway flow, intracranial
         or  semirecumbent  up  to  an  angle  of  approximately  60   pressures and expired carbon dioxide. 30
         degrees. 28
                                                              BLOOD PRESSURE MONITORING
         Zeroing  the  transducer  system  to  atmospheric  pressure
         (calibration  of  the  system)  is  achieved  by  turning  the   Indirect and direct means of monitoring blood pressure
         3-way stopcock nearest to the transducer open to the air,   are widely used in critical care units. These are outlined
         and  closing  it  to  the  patient  and  the  flush  system.  The   in more detail below.
   214   215   216   217   218   219   220   221   222   223   224