Page 298 - ACCCN's Critical Care Nursing
P. 298

Cardiac Rhythm Assessment and Management  275

                                                                  ●  identification of return of spontaneous rhythm
               BOX 11.4  Oversensing: causes and                  ●  assessment of haemodynamic adequacy during both
               management                                            paced and spontaneous rhythms (BP, CO, perfusion,
                                                                     symptoms)
               Causes:                                            ●  strip  documentation  of  rhythm  6-hourly  and  daily
               ●  Muscle potentials other than QRS complexes:        12-lead ECG
                  ●  Cardiac: T waves, U waves, P waves           ●  daily chest X-ray to confirm position of wire/absence
                  ●  Non-cardiac:  shivering,  fasciculations,  seizure  activity,   of complications
                    any skeletal muscle movement                  ●  checking and tightening of all connections (leads to
               ●  External electrical interference:                  bridging cable, bridging cable to pulse generator) at
                  ●  Electrocautery, TENS machines                   commencement of shift and during all pacing adverse
                  ●  Electrical devices (rare)                       events
               ●  Movement of the connecting pins at the connection to the   ●  confirmation of battery status each shift
                  pulse generator (common).                       ●  performance of pacemaker threshold assessment each
                                                                     shift or daily.
               Management:
               ●  Reduce sensitivity (turn sensitivity to higher number)
               ●  Consider disabling the sensitivity altogether (i.e. asynchro-  Protection Against Microshock
                  nous, VOO, AOO, DOO mode)                       Patients with temporary pacemakers require microshock
               ●  Consider  reversing  the  polarity  of  the  wires  (positive  to   protection. Normally, small electrical stimuli (e.g. static
                  negative)                                       electricity  applied  to  the  body)  dissipate  through  body
               ●  Remove  the  source  of  interference  where  it  can  be   tissues and never reach sufficient current density at the
                  identified.                                     heart  to  produce  arrhythmias.  However,  pacing  wires
                                                                  provide a direct route to the heart, so that even minor
                                                                  electrical sources may achieve sufficient current density at
                                                                  the heart to precipitate arrhythmias. Protection strategies
                                                                  include nursing patients in body- and cardiac-protected
             rhythm.  Electromagnetic  interference  resulting  in  over-  areas, insulating external connector pins when pacing is
             sensing  may  arise  from  a  variety  of  causes,  originating   not  in  use,  and  using  rubber  gloves  at  all  times  when
             from the patient (muscle movement) or external sources   handling pacing wires. 70
             (devices). The sources of oversensing 22,51,70,71  may be dif-
             ficult  to  establish  clinically  but  should  be  sought  and
             corrected where able. Causes and management of over-  Battery Depletion in a Temporary Pacemaker
             sensing are detailed in Box 11.4.                    A standard 9V battery might be expected to power a tem-
                                                                  porary pacemaker for up to a week, although this is vari-
             An important distinction must be made between failure   able depending on the device, the mode, rate, and output
             to  pace  and  oversensing  (see  Figures  11.44  and  11.45).   settings, the percentage paced, and the impedance of the
             In both complications the pacing spikes do not appear   pacing leads. Additionally it is often not known whether
             when  expected  and  may  therefore  be  indistinguishable   a  new  battery  was  inserted  into  the  pacemaker  at  the
             from  each  other.  Clearly  the  management  of  the  two   commencement of treatment for the current patient.
             complications  is  different,  and  so  prompt,  accurate
             differentiation  is  important  to  ensure  appropriate   Temporary pacemakers provide indications of depleting
             management.                                          battery status; these are usually displayed when there is
                                                                  less than 24 hours of battery life remaining: (a) flashing
             NURSING PRACTICE                                     battery icons may appear on the digital screens of newer
             Care, monitoring and management of the patient and the   generation  devices;  (b)  on  both  new  and  older  non-
             pacing system largely fall to the nursing staff of critical   digital screen devices, the pacemakers will stop supplying
             care  units.  Nurses  must  ensure  ongoing  monitoring  of   power to the flashing sense/pace LEDs. Battery replace-
             pacing performance and the detection of pacing abnor-  ment should be undertaken as soon as reasonably pos-
             malities, the integrity of the pacing system, the avoidance   sible  as  these  indicators  are  not  obvious  until  looked
             of clinical situations or physical changes that may alter   for,  so  some  time  may  have  elapsed  before  detection
             pacing effectiveness, patient safety, and the prevention of   by staff.
             complications.                                       Changing the battery on a temporary device carries the
                                                                  risk of interrupting pacing which may be disastrous in the
             Nursing responsibilities in the care of the patient with a
             pacemaker include:                                   pacemaker-dependent patient. Although the time taken
                                                                  to change a battery may be brief, additional significant
             ●  pacemaker site inspection for inflammation/swelling/  time may be lost if the device ‘powers down’ during the
                haematoma                                         battery change. It is worth noting, however, that tempo-
             ●  avoidance  of  hip  flexion  and  rest  in  bed  if  femoral   rary  pacemakers  carry  a  small  stored  charge  which  is
                insertion                                         enough to sustain pacing for about 10 seconds. If a well-
             ●  vital  signs,  circulatory  observations,  etc.  at  intervals   rehearsed procedure is undertaken, battery change can be
                appropriate to the overall patient context        performed without interrupting pacing for even a single
             ●  confirmation of capture and sensing               beat. An understanding of the behaviour of the device in
   293   294   295   296   297   298   299   300   301   302   303