Page 299 - ACCCN's Critical Care Nursing
P. 299

276  P R I N C I P L E S   A N D   P R A C T I C E   O F   C R I T I C A L   C A R E


                                        1.  Store current values in memory (for devices with memory). If rhythm difficulty is encountered
                                             during testing, these original settings can be immediately re-established by depressing the
                                             stored values mode selector.
                                        2.  Test for underlying rhythm. Gradually decrease rate in 10 beat/min steps until evidence of
                                             underlying rhythm (ULR) emerges:
                                             (a) If ULR present, observe whether sensing is now occurring.
                                             (b) If still pacing at 50/min (no emergence of ULR), return to initial settings; do not continue to
                                                  test sensitivity or output thresholds.
                                             (c) Document attempt and ULR less than 50/min.
                                        If underlying rhythm is haemodynamically acceptable, continue.
                                        3.  Test sensitivity threshold. Having confirmed haemodynamically adequate underlying rhythm:
                                             (a) Turn pacing rate to half the patient’s rate.
                                             (b) Turn output to minimum (not off).
                                        (NB: Sensitivity testing requires that failure to sense is created for a brief period, so steps 3a
                                        and 3b are designed to minimise danger of arrhythmias.)
                                        While observing the sense indicator on the pacemaker:
                                             •    decrease sensitivity (increasing the number) until failure to sense (the sense indicator
                                                  stops flashing — pacing indicator will now be flashing);
                                             •    increase sensitivity (decreasing the number) until sensing resumes;
                                             •    note the value at which sensing returns — this is the threshold value for sensing; and
                                             •    set sensitivity to half this value minus 1 mV.
                                        4.  Test output threshold (continuing from step 3 above the pacing will now be set at a low rate,
                                             and at minimum output):
                                             •    Increase the pacing rate to 10 greater than underlying rhythm.
                                             While watching the monitor:
                                             •    gradually increase the output until capture is achieved;
                                             •    note the value at which capture occurs — this is the threshold value; and
         FIGURE  11.46  Routine  temporary        •    set output to double this value plus 1 mA.
         pacemaker  testing  protocol:  underly-  5.  Store new values in memory and document settings.
         ing  rhythm,  output  and  sensitivity
         threshold test.

         use  should  be  established  before  undertaking  battery   undertaken and to report any sensations of lightheaded-
         replacement.                                         ness, dyspnoea or other discomfort.
         Pacemaker Function Testing                           Pacemaker testing in the unstable
         Routine pacemaker performance checks should be under-  pacemaker-dependent patient
         taken  regularly  in  the  patient  with  a  temporary  pace-  Greater caution must be applied in the testing of pace-
         maker.  Temporary  pacing  leads  and  wires  are  prone  to   maker  functions  if  the  patient  has  marked  haemody-
         movement and therefore to sensing and capture thresh-  namic instability or has little or no underlying rhythm.
         old variation. Variations may also be marked when there   It  is  common  for  pacemaker  testing  to  be  avoided
         is myocardial, biochemical and haemodynamic volatility   altogether  in  such  circumstances  although  this  may  be
         as often seen in the critically ill patient. Pacemaker tests   misguided.  Routine  testing  of  pacemaker  function  as
         are performed to reveal the return of underlying rhythm   described in Figure 11.46 may not be suitable, but testing
         which may be being concealed by pacing, and to measure   for  underlying  rhythm,  and  some  level  of  testing  of
         thresholds for both capture and sensing, as these values   capture threshold so as to be confident of safety margins
         typically change with time and in response to changing   is beneficial. For the patient with haemodynamic insta-
         myocardial  responsiveness. 54,58,62,68   Regular  checking   bility and/or inotrope use, testing for underlying rhythm
         allows  detection  of  threshold  changes,  and  setting  of   becomes of even greater important as pacing may either
         sensing and output safety margins, in order to minimise   prevent or conceal the return of sinus rhythm capability,
         the  development  of  acute  failure  to  capture  or  failure    and cardiac output may be as much as 50% greater with
         to sense.
                                                              the atrial kick of sinus rhythm than during pacing (see
         The  practices  employed  to  test  temporary  pacemakers   Figure 11.47). It may take several seconds for the sinus
         vary widely across Australia, as do attitudes to whether   node to ‘warm up’ and express itself, so decrease the rate
         this may or may not be undertaken by nurses. The sample   gradually and only to reasonable levels (sinus rates of less
         protocol  shown  in  Figure  11.46  provides  an  organised   than  50  are  unlikely  to  be  beneficial).  Be  sure  to  gain
         approach to testing during which safety has been empha-  agreement  from  the  multi-disciplinary  team  before
         sised. Because of the varying attitudes to nursing respon-  undertaking testing in this context.
         sibilities, the use of this approach should be ratified at
         individual institutions before use.                  Threshold testing in the pacemaker-dependent patient is
                                                              also contentious as loss of capture during testing may be
         Testing  pacemaker  thresholds  is  performed  daily  or  on   poorly tolerated. If the capture threshold is not measured,
         each  shift,  but  not  if  the  patient  is  unstable,  using  the   however,  a  rising  threshold  and  loss  of  safety  margins
         steps described in Figure 11.46. The test should be carried   cannot  be  identified,  and  may  only  become  apparent
         out promptly, with attention to avoiding undue bradycar-  upon development of acute failure to capture, possibly
         dia  or  periods  of  asynchronous  pacing.  The  patient     with outputs already set to maximum and therefore no
         should  be  advised  that  pacemaker  assessment  is  being   scope for recovering capture. An alternative approach to
   294   295   296   297   298   299   300   301   302   303   304